A Novel Dual Energy CT-Based Attenuation Correction Method in PET/CT Systems: A Phantom Study

 

Behnoosh Teimourian1, 2, Mohammad Reza Ay2,3,4, Mojtaba Shamsaei Zafarghandi1, Hossein Ghadiri2, 5

 1Faculty of Nuclear Engineering and Physics, Amir Kabir University of Technology, 2Research Center for Science and Technology in Medicine,3Department of Medical Physics and Biomedical Engineering,  4Research Institute for Nuclear Medicine, Tehran University of Medical Sciences, Tehran, Iran

5Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran

Corresponding Author:
 

Dr Mohammad Reza Ay, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.

E-mail: mohammadreza_ay@tums.ac.ir

 

Received:October 17,2009
Accepted: November 24,2009
Available online:December 20,2009
Abstract:

In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-energy technique with only a single energy CT scan. The implementation was done by CT scans of RANDO phantom at tube voltages of 80 kVP and 140 kVP. The acquired data was used to obtain conversion curves (which scale CT numbers at different kVP to each other), in three regions including lung tissue (HU<-100), soft tissue (-100<HU<200) and bone tissue (HU>200) for the combination of 80 kVP /140 kVP. Therefore, with having the CT image in one energy, we generate the CT image at the second energy (from now we call it virtual dual-energy technique) using these kVP conversion curves. The attenuation map at 511 keV was generated using bilinear (the most commonly used method in commercially available PET/CT scanners), real dual-energy and virtual dual-energy technique in a polyethylene phantom. In the phantom study, the created attenuation map using mentioned methods are compared to the theoretical values calculated using XCOM cross section library. The results in the phantom data show 10.1 %, 4.2 % and 4.3 % errors for bilinear, dual-energy and virtual dual-energy techniques respectively. Further evaluation using a larger patient data is underway to evaluate the potential of the technique in a clinical setting.

 

Keywords:
PET/CT ،  Dual Energy CT ،  Attenuation Correction