Volume 29, Issue 3 (10-2016)                   jdm 2016, 29(3): 159-168 | Back to browse issues page

XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Pirjamalineisiani A, Sarafbidabad M, Jamshidi N. Biomechanical comparison between pure Ti and Zr -2.5%Nb dental implants after implantation by finite element analysis (FEA). jdm 2016; 29 (3) :159-168
URL: http://jdm.tums.ac.ir/article-1-5559-en.html
1- Researcher, Department of Biomedical Engineering, Faculty of Engineering, University of Isfahan, Isfahan, Iran
2- Assistant Professor, Department of Biomedical Engineering, Faculty of Engineering, University of Isfahan, Isfahan, Iran
Abstract:   (6529 Views)

Background and Aims: Improving dental implantation conditions in order to reduce the failure is always desirable for researchers. The aim of this study was to compare two different materials of dental implants from the viewpoint of biomechanical effect after placement and loading in the mandible.

Materials and Methods: A 3D model of mandible was designed in the MIMICS 10.01 software. Then, by using the obtained model, the end part of the mandible was designed in CATIA V5 software and a drilling operation was performed on the cortical bone of mandible by finite element analysis simulation method in DEFORM-3D V6.1 software. Thereafter, the 3D model of created hole was extract from the drilled site and an ITI dental implant model designed in the CATIA V5 software, was placed in the hole. The space remained between the implant and cavity was considered as a newly-formed cortical bone in drilled site after 6 weeks of dental implant placement and the mechanical properties of newly-formed bone were entered to DEFORM-3D V6.1 software. Then, a load was applied on the top surfaces of two dental implant models with the materials titanium and zirconium-2.5% niobium.

Results: The emerged volume changes in newly formed cortical bone around Ti and Zr-2.5 Nb dental implants were measured 0.238 and 0.242 percent, respectively. Furthermore, micro-motion of Ti and Zr-2.5 Nb dental implants were measured 0.00514 and 0.00538 mm, respectively.

Conclusion: The results of this study showed that Ti dental implant created better conditions than Zr-2.5 Nb dental implant in the mandible.

Full-Text [PDF 1046 kb]   (4814 Downloads)    
Type of Study: Research | Subject: general
Received: 2017/01/29 | Accepted: 2017/01/29 | Published: 2017/01/29

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and Permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 , Tehran University of Medical Sciences, CC BY-NC 4.0

Designed & Developed by: Yektaweb