Fahimeh Mokhtari, Amir Mohamad Mahabadi, Naieeme Ghafarian, Alireza Ghate, Alireza Razavi,
Volume 25, Issue 1 (7 2012)
Abstract

Background and Aims: Working length determination and remaining this length is of great importance in root canal therapy. Recently, electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the sensitivity of Root ZX and Novapex electronic apex locators with radiography in a clinical study.

Materials and Methods: In this study, 73 single-canalled teeth of the patients referred to the endodontic department of the Yazd Dental School were evaluated. The access cavity was prepared and working length of the root canals were measured by using the apex locators: Root ZX and Novapex. The file was placed in the root canal and a periapical radiograph was taken using parallel technique. The working lengths obtained by apex locators were recorded and compared with those of radiographs. The data were analyzed by Pearson's correlation coefficient, Wilcoxon test and paired t-test.

Results: The exact measurement without any fault was 46.6% for Root ZX and 20.5% for Novapex compared to radiography. In the range of 0.5 mm fault, the results were 91.8% for Root ZX and 64.4% for Novapex compared to radiographic measurements. The difference between the average of the measured lengths by Root ZX and radiography was not significant (P=0.17). On the other hand, the results showed significant difference between the average of the measured lengths by Novapex and radiography (P=0.017). The difference between two apex locators was not significant (P=0.061).

Conclusion: Root ZX and Novapex with high accuracy are useful for determining working length of the root canals in pregnant patients with special anatomic conditions and patients with muscle disharmony. Therefore, the use of these apex locators combined with radiography is recommended in root canal treatment.


Jalil Modaresi, Amir Reza Heshmat Mohajer, Hossein Aghili, Morteza Khorshidi, Faramarz Rostami Chavoshlo, Amirmohamad Mahabadi,
Volume 25, Issue 3 (1 2012)
Abstract

Background and Aims: Electric pulp testing is used for diagnosis of the pulp status. This test is technique sensitive and hence may elicit positive or negative false response in case of inaccurate use. The optimal site for placement of the probe tip has not been determined. The aim of this study was to evaluate the effect of pulp tester probe placement site on the response of maxillary teeth.
Materials and Methods: A total of 378 teeth (126 central incisors, 126 lateral incisors and 126 canines) in 67 voluntary 20-35 year-old patients were selected. Three sites on each tooth (incisal edge, labial and lingual surfaces) were tested with an electrical pulp tester and responses were recorded. Data were analyzed by Repeated Measure ANOVA test.
Results: The central teeth showed significantly lower sensation threshold than lateral and canine teeth (P<0.001). The incisal edge of tooth were significantly more sensitive compared to labial and lingual surfaces (P=0.008).
Conclusion: This study showed that the optimum site for placement of pulp tester probe was incisal edge.