Volume 25, Issue 3 (10-2019)                   Back to this Issue | Back to browse issues page

XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Sayadi L, Seylani K, Akbari Sarruei M, Faghihzadeh E. Physiologic monitor alarm status and nurses’ alarm fatigue in coronary care units . Journal of Hayat 2019; 25 (3) :342-355
URL: http://hayat.tums.ac.ir/article-1-3143-en.html
1- Dept. of Medical Surgical Nursing, School of Nursing and Midwifery, Tehran University of Medical Sciences, Tehran, Iran; Nursing and Midwifery Care Research Center, School of Nursing and Midwifery, Tehran University of Medical Sciences, Tehran, Iran
2- Dept. of Critical Care Nursing and Management, School of Nursing and Midwifery, Tehran University of Medical Sciences, Tehran, Iran
3- Dept. of Medical Surgical Nursing, School of Nursing and Midwifery, Tehran University of Medical Sciences, Tehran, Iran , m.valiasr12@gmail.com
4- Dept. of Biostatistics, School of Paramedical Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran
Abstract:   (4136 Views)
Background & Aim: Patient monitoring in cardiac intensive care unit (CICU) is considered to be non-invasive and safe, but it can cause unintentional catastrophic consequences due to the alarm fatigue. The aim of the study was to determine the status of monitoring system alarms and nurses’ alarm fatigue.
Methods & Materials: This descriptive, cross-sectional study was carried out in cardiac intensive care units of a hospital in Fars province from January to March 2019. The alarm status of 24 monitoring devices was observed for 100 hours in 100 patients under the physiological monitoring. Patients under monitoring were included in the study using a continuous and convenience sampling. Also, 62 nurses working in five CICUs of this hospital completed the alarm fatigue questionnaire (through census). The data obtained from the observation and the questionnaires were analyzed by the SPSS software version 16 using descriptive statistics.
Results: The auditory alarms of monitoring devices were turned off and only visual alarms could be recorded. In the visual alarms recorded during 100 hours, 131 alarms (53.47%) were technical, followed by the false, nuisance, and ultimately real or actual alarms with rates of 40 (16.33%), 38 (15.51%) and 36 (14.69%), respectively. The mean (and standard deviation) score of alarm fatigue among nurses was 21.04 (7.52), and the alarm fatigue score for %50 of nurses was higher than the average score.
Conclusion: Turning off the auditory alarms can lead to serious risks to patients. Also, a high proportion of technical, nuisance, and false alarms can cause alarm fatigue in nurses. Adopting solutions in accordance with standard guidelines and checking physiological monitoring devices in hospitals are necessary in order to reduce false, nuisance, and technical alarms. Failure to pay attention to these issues leads to the alarm fatigue among nurses, which itself results in numerous consequences such as compromising the patients’ safety.
 
Full-Text [PDF 255 kb]   (1985 Downloads)    
Type of Study: Research | Subject: Nursing Care
Published: 2019/12/15 | ePublished: 2019/12/15

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 , Tehran University of Medical Sciences, CC BY-NC 4.0

Designed & Developed by : Yektaweb